Support
for equipment was provided by the National Science Foundation,
Division of Undergraduate Education (UNDERGRADUATE INSTRUMENTATION
AND LAB IMPROVEMENT GRANT #9451930) to The Department of Biology,
Wartburg College.
Operation
of the S-2460N SEM
This
primer attempts to cover the procedures to follow for using the
Hitachi SEM from the time you walk in the room until you finish
for the day. Any and all comments or corrections or complements
or complaints are welcome and will be used to make this a better
tool.

Fig.1.
Hitachi Scanning Electron Microscope Model S-2460N
A -
Power Switch E - Digital Video Processor
B - Standard Stage F - Control Panel
C - Flicker-Free Image CRT G - Keyboard
D - Raw Image CRT H Polaroid camera
Figure
1 shows the user interface of the Hitachi S-2460N. In addition
to this unit, there are three vacuum pumps that run 24-7-52 (24
hours and day-7 days a week-52 weeks a year) in the back room.
Normally, when you attempt to use the SEM, the display power will
be off (A), the column and sample stage will be under a high vacuum
(B), and the vacuum pumps should be running (not shown).
PROCEDURES
1. Enter
your name, date, and the starting beam time into the record book.
2. Toggle ON the Display Power by your left knee (bottom left
Fig.1 & Fig. 2). The EVAC POWER on the left must always be
on as it controls the vacuum pumps which stay on all the time.
In case of a power outage the SEM shuts itself down and powers
up the vacuum pumps when the power comes back on.


Figure
3. The sample stage and column.
A
- Stage X Control (-40 to 40 mm)
B - Stage T Control (-20 to 90 degrees)
C - Stage R Control (360 degrees)
D - Stage Y Control (-20 to 20 mm)
E - Stage Z Control (5 to 35 mm)
F Infrared video camera (view of sample chamber)
G- Robinson detector (NSEM mode).
3.
Make sure the stage (Fig. 3) is at the proper exchange
settings. Especially knob E where the white "tabs"
must be lined up!!! This ensures that the specimen stub is directly
beneath the pole piece (where the beam comes out) AND it
also ensures the sample is low enough not to hit the detector.
4. Since
the vacuum is always on, we must allow air into the column in
order to remove the sample holder. Press the AIR button (Fig.
4) and the vacuum will be released. Wait until the door can be
opened (about 2 minutes).

Figure
4. Evacuating System Control Panel. The AIR/EVAC button is used
to start/end the vacuum in the chamber. In the SEM mode, the meter
should read ~10 Pa and NSEM mode ~20 Pa. The lights (HIGH/SE/BSE)
will indicate when the vacuum is sufficient to saturate the filament.
5. At
this point you should turn on the IR camera (Fig. 5, F in Fig. 3) to view the inside of the chamber and check out the
status of the stub holder and whether samples are in the scope.
Press the white button below. Each click will switch from chamber
view to SEM mode.
. ë
Fig. 5. Chamberscope IR camera located above Keyboard G.
6.
If chamber looks clear, carefully open the STAGE DOOR and
place a stub holder (Fig. 6) in the stage mount and check height
of specimen so that it clears the pole piece and back scatter
detector as the door is closed. If you set the stage height
correctly in STEP 3, there should be no problem. ***Use the metal
height plate to be sure***

Figure
6. The five sample holder we use quite a bit. Allows for five
samples to be viewed. Knob C the stage R Control (360 degrees)
in Figure 3 rotates the sample under the beam.
BE SURE
THE FILAMENT KNOB IS SET TO ZERO PRIOR TO NEXT STEP !!!

Figure
7. The filament potentiometer. It is located under the "Secret
Panel". Knob at the left is used to turn up the filament
current to a saturated (optimum) level. The level is measured
with the emission current meter and the raster image which comes
up on the Raw Image CRT (Fig. 1). It must be at zero when you
start. Generally if you set it to 1 oclock you will be very
close to the optimum (saturated) current level.
7.
Be sure white chamber scope is set to SEM mode so you can see
the screen. Use F2 on keyboard to choose your mode
SE/NSEM (Fig. 8). If you chose NSEM you must insert the Robinson
detector into the chamber. Carefully release the pin and the vacuum
will pull in the detector quickly!!! If using the SE mode, the
Robinson detector should be out. It is tough but you can pull
it out against the vacuum. Then use F8 to set the beam
current to the range you want.

Figure
8. Keyboard. "F" keys across the top control functions
on the scope.
F1
after a photo, click F1 to reset image back to "live"
video.
F2 sets the NSEM/SE mode when starting the vacuum.
F10 Sets the beam current.
8. Slide
the stage door closed, press EVAC button. Wait for vacuum
lights to change from low to high and red HIGH Kv lamp
to burn steadily (2-3 minutes). The SE or NSEM light (Fig. 4)
will glow also. When the READY LIGHT stops blinking, choose
desired Accelerating Voltage by pressing the "up/down" buttons (Fig. 9) then turn on ACC. Voltage by pressing "ON/OFF" button.

Figure
9. The ready light indicates conditions are such that you can
begin to saturate the filament (push the ON/OFF switch). One must
set the accelerating voltage (UP/DOWN) prior to saturating the
filament. As you turn up the current to the filament, a color
bar will form on the emission current meter .
10.
Saturate the filament using the static waveform. Push waveform
button twice (Fig. 10, upper right corner) slowly turn filament
knob clockwise (Fig. 7 UNDER THE SECRET COMPARTMENT) until the
waveform reaches its highest level (post HV button must be turned
on for SE mode). Once set, you can view the sample image by pressing
one of the span speed buttons (Fig. 10, top row). We usually use
the computer enhanced "live" mode (leftmost button).
Click it twice to get computer smoothed image.

Figure
10. Upper panel shows the scanning speed buttons on left.
The open box is "real-time" video [The computer takes
the scanned image and makes it appear as a steady "TV"
image]. The middle ones: you see the scan across the image, last
one: this makes a small box on screen to help with focus.
Waveform
buttons in upper right corner are used to obtain waveform
image on the Raw Image CRT (Fig. 1) so you can saturate the filament.
Lower
panel right shows the focus controls. THEY ARE MISLABELED:
The bottom two buttons allow you to coarse focus (up and down)
and the knob is the fine focus.
11.
Set the brightness and contrast to AUTO and then Press ABC (Fig. 11). This will automatically (or at least try to) set
these parameters. You might need to go to manual.


Figure
11. Upper - brightness and contrast knobs. In the AUTO mode pressing
the ABC will allow the computer to best determine the best settings
(sometimes it gets worse than your settings). In manual setting
you spin the knobs with your finger to adjust the image to your
liking. Lower Stigmator knobs fix astigmatism. At high
magnification and in the SE mode, the image may be "out of
round" so you need to adjust it. AUTO works OK but generally
manual settings gives a better image.
12.
Set desired working distance with knob E (Fig. 3). Working
distance can be monitored by pressing the down arrows focus button
once and it will read out at the magnification LED.
Minimum
W.D. : BSE 10mm with the Robinson detector in
13. Focus image (Fig. 10). One can set the focus to AUTO and
press either button and the computer will try to focus on your
sample OR one can focus manually with coarse buttons and fine
knob. Increase/decrease the magnification with the magnification
buttons on far right of Panel F (see Fig. 12. You will do this
a lot!!!
14.
Move the sample using the controls on the STAGE (Fig 3) or for
small small movement, the IMAGE SHIFT (Fig. 9). At his point you
will likely "PLAY" with the brightness/contrast/focus/magnification,
etc
to get the best picture. This is where the ART of SEM
comes in. Bottom line: Focus your sample at the highest mag
you can and then capture image at lower mag
.
Finally
you have the perfect image !!!!
15.
Shoot a picture. Pick a scan speed (1/2/3) in the PHOTO system
(Fig. 12) to view the best image of your sample. Once you like
the image, capture the image by pressing the PHOTO button.
A fine image will slowly appear as the computer collects lots-o-data
which produces a high res image.

Figure
12. The upper panel shows the photo system. After getting the
best image possible in the "real-time" mode one then
sets the time of scanning with the speed switch and then depresses
the PHOTO button which starts the very slow scanning needed to
develop the best still image. The lower panel shows the magnification
buttons. Use up and down zoom in and out
.
16.
Print a picture with the VIDEO PROCESSOR.

A
- Power Switch
B Processor controls
C - Print Switch
D Printer paper control
Fig.
13. Digital Video Processor. Move switch D down to print position
when ready to print. Press C or use remote print switch to print
the picture after it is captured using the PHOTO button above
the keyboard (Figure 9). This is a digital printer so the "film"
(paper) is NOT ruined when exposed to light.
17.
Once you have printed the picture, press F1 to unfreeze
the image so you can look around some more, take some pictures,
whatever
.
To
exchange specimens:
- Turn
down the filament.
- Reset
stage to proper exchange settings.
- Press
the AIR button and wait until door opens.
- Remove
samples. Close the door and place new samples on the holder.
Open the door, insert holder.
- Close
door and press EVAC button.
- Go
to step 8 above.
To
take a break:
Turn
down the filament, leave everything else as is
turn up filament
when you come back.
To
end your session:
- Turn
down the filament.
- Reset
stage to proper exchange settings.
- Press
the AIR button and wait until door opens.
- Remove
samples.
- Close
door and press EVAC button. If ending session, leave
microscope in the SE mode (25 Pa, high vacuum).
- Shut
off DISPLAY POWER.
- Log
out in the record book.

Figure
14. Single sample stud holder along with black conductive stub
two-sided tape. Simply peel the circles off the paper and apply
to the stub. When your sample is ready, peel off the upper black
cover and apply sample to the stub. Process the sample and view.
Photographing
Condition Setting
- Select
a desired DIRECT PHOTO exposure time by the corresponding SPEED selector switch.
1: 40 sec/frame
2: 80 sec/frame (used ordinarily)
3: 320sec/frame (for X-ray microarea analysis or the like)
- Only
the left or right half of an image can be exposed by turning
the TWIN switch to L or R. Utilize this
function according to your microscopic purpose.
Direct
photo procedure
- Select
a visual field, and carry out focusing and astigmatism correction.
- Optimize
image brightness and contrast.
- Press
the PHOTO switch. Image disappears from the flicker-free
image CRT and is newly formed on the screen. Scan is performed
downward on the raw image CRT.. While photographing, the selected
one of the exposure lamps 1,2, and 3 flickers. When scan
reaches the lowermost line, the buzzer beeps to notify the end
of image recording. On the flicker-free image CRT, the recorded
image is frozen and the scan speed before activation of the
PHOTO mode returns on the raw image CRT. For cancellation of
this photographic recording midway, press the PHOTO switch
again.
Notes: 1. IN the course of recording, all switches except for ACC
VOLTAGE ON/OFF will not function.
4. Once
scan has covered two thirds of the screen after start of recording,
its cancellation cannot be accepted.